商品介紹生技相關儀器GL220系列 生物冷光儀

GL220系列 生物冷光儀



















Kinetic Measurement Mode

- Time course of luminescence intensity is measured in the kinetic measurement.

- Measurements for flash luminescence system in which the luminous phenomenon disappears instantaneously.

- Measurements in which the slope of the luminescence curve right after reagent addition is important.

- GL-220S performs the kinetic measurement effectively with a syringe injection port.








RLU Measurement Mode

- Measures the RLU (Relative Light Units) per unit time.

- Basic mode to measure the integrated value of luminescence.

- Creation of calibration curves

- Calculation of concentration

 

Dual Luciferase Assay Mode

- Corresponding to the dual luciferase assay that is one of the reporter gene assays.

- Performs complex analytical calculation with built-in software.

- Protocol for the dual-luciferase assay

- Automatic calculation of specific activity value



Specificationse


 


GL-220

Standard type

GL-220A

High-sensitivity type

GL-220S

High-sensitivity type

with a syringe injection port

Sensitivity

5×10-12

mole ATP

5×10-13

mole ATP

Measurement

Modes

1. Kinetic Measurement Mode

2. RLU Measurement Mode

3. Dual Luciferase Assay Mode

Mycoplasma detection mode (option)

Measurement

time

1 ~ 99 sec. (Settable with the unit of 1 sec.)

Settings

Measurement mode, Measurement time, Number of measurements,

Number of samples, and etc.

Output

‘Date & time’, ‘Measurement mode’, ‘Measurement conditions’, and ‘Calculation values’ can be output to the serial port.

Sample volume

100 ~ 500 μL

Dimension

160(W)×240(D)×104(H) mm (for GL-220, GL-220A)

160(W)×240(D)×112(H) mm (for GL-220S) (not including protrusions)

Weight

1.5 kg

1.6 kg